NEL Frequency Controls, Inc.
About NELAbout Nel - Japanese

What's New at NEL Frequency Controls

Join Our Email List

TESTING PHASE NOISE OF ULTRA LOW PHASE NOISE OCXO – CHALLENGES AND SOLUTIONS

Roman Boroditsky, Jorge Gomez
NEL Frequency Controls
Burlington, WI USA
romanb@nelfc.com, gomezj@nelfc.com
Updated August 23, 2021

Abstract—Modern instrumentation equipment, commercial & defense communication systems, and radar equipment require excellent stability, extremely low phase noise frequency sources. Testing devices that exhibit phase noise of -120 dBc/Hz at 1 Hz offset from the carrier and better than -180 dBc/Hz on the noise floor is a challenge using existing test equipment and methods. It is especially pertinent to production environment, where measurement time and accuracy of each measurement becomes critical. The purpose of this work was to investigate different test methods, evaluate different phase noise measurement equipment, and find acceptable solutions for both low frequency (around 10 MHz) Ultra Low Phase Noise (ULPN) reference, and HF/UHF ULPN OCXO. Several test methods and test instruments were investigated. There's no "one size fits all" solution, but for each frequency range the optimum solutions were proposed.

Introduction, Challenges
There are several criteria that define acceptability of the test methods and equipment used in the measurements of phase noise of ultra low noise (ULPN) OCXO. These are:
1. Accuracy
2. Repeatability
3. Speed of test
4. Ease of use
5. Range
6. Cost
7. Ease of data retrieval

The value of each characteristic also depends on whether the test is intended for use in production environment, or in the lab. The "two oscillator method" [1] was ruled out, since it does not satisfy the first four criteria very well, and therefore only methods and equipment, employing cross-correlation technique were considered. [1], [4]. The challenges for lower frequency devices (around 10.000 MHz) were :

  • Achieving measurement accuracy in a reasonable amount of time at offset points close to the carrier
  • Reaching the noise floor farther away from the carrier.

For higher frequency ULPN OCXO, specifically at 100.000 MHz carrier frequency, RF pollution in the environment, and EMI added further challenges.

For devices, which employ a phase lock loop (PLL) technique for attaining higher frequency the interaction of the device PLL and test instrument's PLLs made it even more difficult to achieve accurate results.

I. LOW FREQUENCY ULPN OCXO

The obvious choice of equipment for testing low frequency OCXOs was Symmetricomm's 5120A Phase Noise Test Set with option 01 (internal reference) [3], which uses cross-correlation technique. It satisfies the majority of the established criteria, and does not require additional references, nor calibration of the device under test (DUT) on exact frequency. The phase noise plot of 10.000 MHz ULPN OCXO is shown on Fig.1.


Fig.1. 10 MHz ULPN OCXO on 5120A – 01

The only concern was the resolution (noise floor) of the measurement at large offsets from the carrier (the noise floor of the device). The instrument's specification calls (conservatively) for -168 dBc/Hz floor at offsets greater than 10 KHz. The theoretical expectations were closer to -175 dBc/Hz at 100 KHz offsets and beyond. We decided to verify our expectations with a Noise XT Dual Core Noise Test Set (DCNTS) [2], [4]. It is a considerably more difficult type of test. It requires two references with similar performance as the DUT (the better the references' performance – the faster the test), the references must have voltage control (ability to change frequency with the change of the voltage on the control terminal), and be calibrated exactly on the frequency of DUT. The test result of the same device is shown in Fig.2.


Fig.2. 10.000 MHz ULPN OCXO on DCNTS

As expected, the measured phase noise far from the carrier was improved by 3 – 5 dB, while discrepancies in results from a few KHz offset in were not very noticeable. Realizing that the phase noise on the noise floor is determined by the circuit, and for given design unit to unit variations are negligible, the instrument of choice still would still be 5120A-01, however, during the design qualification stage, the verification of the noise floor values on the DCNTS is required.

II. 100.000 MHZ ULPN OCXO

There's no ideal option to satisfy all established criteria. We started with presumably the most accurate and fastest instrument – the DCNTS from Noise XT. The first results were surprising and the plot is shown in Fig. 3.


Fig. 3. 100 MHz ULPN OCXO with "bumps" caused by RF pollution.

After a lengthy and thorough investigation it turned out that the spurious "bumps" at 100 KHz, 700 KHz, and 900 KHz offsets from the carrier were caused by mixing the broadcast signals from the local FM radio stations (at 100.1 MHz, 100.7 MHz, 99.1 MHz) with 100 MHz OCXO signal. While testing good, but not ULPN 100 MHz devices, (which have noise floor at -170 dBc/Hz or higher), this phenomenon is hardly noticeable. Therefore this challenge is specific to ULPN OCXO with a phase noise floor around -180 dBc/Hz, particularly running at 100 MHz, especially in urban areas, where the power from the FM stations is quite high. If the frequency is far away from the broadcasting FM band, this phenomenon doesn't occure. In order to improve this condition we took several measures:

  • All power supply cables were shielded and connected to the test fixtures via SMA connectors.
  • Large value electrolytic capacitors were placed for decoupling power supply cords
  • We built RF shielding cage and placed all test equipment, including power supplies and test fixtures inside the cage as shown on Fig. 4.


Fig. 4. RF shielded test station.

We tested the unit whose phase noise plot is shown in Fig. 3. with the improved setup. The resultant plot is shown in Fig 5. We also included the plot (as a courtesy of Guillaume De Giovanni, Noise XT) with data taken on a batch of ULPN OCXOs in a benign (RF pollution-wise) environment. The test instrument is the same DCNTS with the "smooth" function used. It's shown on Fig. 6.


Fig. 5. 100.000 MHz ULPN OCXO phase noise tested with RF-shielded setup.


Fig. 6. A batch of 100.000 MHZ ULPN OCXO tests in benign environment. Smooth function is used.

From the ease of use standpoint, a very attractive option is Agilent's E5052B Signal Source Analyzer [5]. Also testing was performed on the device, similar in performance to the one shown in Fig. 5. using the E5052B. The instrument utilizes built-in synthesizers and the cross-correlation method. The plot of the experiment with varying numbers of cross correlations and averages is shown in Fig. 7. As shown in the plot, the instrument can resolve the performance of a ULPN device with sufficient number of cross-correlations, but unfortunately it takes an enormous amount of time to do so. One would have to run the test for several hours to get the noise floor of the instrument at all offset points low enough to achieve an accurate measurement. That would rule out the use of the instrument in production environment, but it still maybe useful in R&D, since it doesn't require additional references, or calibration of the references and devices (setting on exact frequency), etc. If the performance on the noise floor of the device far from the carrier frequency is the only parameter needed, the measurement time can be significantly reduced.


Fig. 7. 100 MHz ULPN OCXO tested on Agilent E5052B with different number of cross-correlations and averages.

From a cost standpoint, there's one new little instrument made by Swiss company Anapico APPH6000 [6]. It doesn't have as many options and capabilities as the DCNTS system, and is still in the process of improving GUI and firmware, but it tests reasonably fast and can be used successfully in production. Test results are not as presentable as other instruments, but acceptable. Example is shown on Fig. 8.


Fig. 8. 100 MHz ULPN OCXO tested on Anapico APPH6000

III. ULPN OCXO WITH MULTIPLICATION

A. Analog harmonic multiplication
We tested several ULPN OCXOs, which employed analog harmonic multiplication, with references ranging from 10 MHz to 100 MHz and multiplication factors from 2 to 15 on the DCNTS from Noise XT. The instrument handled tests without problems and some of the examples are shown on Fig. 9. through Fig. 12.


Fig. 9. 80 MHz ULPN OCXO, multiplied from 10 MHz


Fig. 10. 750 MHz ULPN OCXO, multiplied from 50 MHz


Fig. 11. 1 GHz ULPN OCXO, multiplied from 100 MHz


Fig. 12. 40 MHz OCXO, multiplied from 10 MHz

We also tested one of those units (100.000 MHz, multiplied from 10.000 MHz) on the E5052B with different number of cross-correlations and averages. The plot is shown on Fig. 13. As when testing ULPN devices with no multiplication, despite the ease of use, the speed of test required to achieve meaningful resolution is not acceptable for a production environment. With 1,000 cross correlations it took over 2.5 hours for the first trace (1 average) to appear.


Fig. 13. 100 MHz ULPN OCXO multiplied from 10 MHz, tested on Agilent E5052B with various number of crosscorrelations and averages

B. ULPN OCXO with PLL multiplication
To achieve "the best of the both worlds" i.e. combine the best close to the carrier phase noise, short term, long term, and environmental stability of the low frequency (10 MHz) ULPN OCXO with the lowest phase noise on the noise floor of high frequency (100 MHz) ULPN OCXO, the latter is locked to the former in a low noise PLL, as it is done in NEL's dual frequency reference module (DFRM). Testing of the 10 MHz output can be done as described in chapter I, without presenting any additional challenges. It is much more difficult to achieve realistic test results on the 100 MHz output while testing with two other DFRMs as references. The interaction of internal PLLs with instrument's PLLs along with necessary filtering on control ports of the modules creates artifacts in the phase noise plot, which make the data mostly unusable. The only solution we found so far is to make to test runs for different offset ranges, with different references, and concatenate them afterwards. The result of this effort is shown on Fig.14. For higher frequency offsets we used nonmultiplied ULPN OCXOs at 100 MHz as references. They were free running with voltage control used by the instrument's PLL. For lower frequency offsets we used a ULPN OCXO at 100 MHz, which were harmonically multiplied from 10 MHz, as references. The splicing point is at 100 Hz, just below the internal PLL loop bandwidth.


Fig. 14. DFRM phase noise. Outputs at 10 MHz and 100 MHz

IV. CONCLUSIONS

Testing phase noise of ULPN OCXOs requires using the cross-correlation technique. Special care must be taken for reduction of RF interference, especially while testing 100 MHz ULPN OCXO in the vicinity of strong FM broadcasting stations. The choices of readily available equipment are limited, and the instruments are suited differently to different types of DUTs and different purposes of test. Our rating of evaluated instruments is presented in Table I. We want to emphasize that this is applicable for ULPN devices only. We had very little experience with the Rohde & Schwartz signal analyzer, but expect that it would rate similarly to E5052. Rating is 1 to 5, five stars is being the best, one - the worst.

TABLE I.

CategoryTest Instrument for testing phase noise ULPN OCXO
DCNTSE5052B5120A-01APPH6000
Accuracy********* (note1)****(note2)****
Speed****** (note3)*********
Ease of use************
Range***********(note4)***
Cost***********
Ease of data retrieval***************

Notes:
1. For E5052B the accuracy is good after very long test time.
2. For 5120A-01, the resolution on the floor is insufficient.
3. To achieve accurate data.
4. Limited to 30 MHz

ACKNOWLEDGMENT
Authors would like to thank Guillaume De Giovanni from Noise XT for support in better understanding DCNTS operation and providing test data, and Dr. Jakub Kucera from Anapico for support and updates.

REFERENCES
[1] F.L. Walls et al. Extending the Range and Accuracy of Phase Noise Measurements. Proceedings of the 42nd Annual Symposium on Frequency Control, 1988.
[2] G. De Giovanni, M. Chomiki. New Phase Measurement Techniques and Ultra Low Noise SAW Oscillator. Proceedings of IEEE International Frequency Control Symposium, 2010, pp. 116 – 118.
[3] Symmetricom. 5120A Data Sheet.
[4] Noise XT. DNTCS manual.
[5] Agilent. E5052B Data Sheet.
[6] Anapico. APPH6000 Data Sheet.

ADDENDUM

Since the paper was written, a couple more capable instruments come into play for phase noise testing of ULPN crystal oscillators. The most notable is Rohde &Schwarz R∓SRFSWP-B61 Low phase noise with cross correlation. The instrument is fast (in most cases), easy to use, covers wide range of input frequencies and offset frequencies. It allows to relatively easy test DFRM and Multi Frequency Reference Module (MFRM) higher frequency outputs phase noise in one shot without concatenation (splicing). Here's an example of the MFRM all outputs obtained without splicing on the same screen.

Another example is 1 GHz ULPN OCXO. The test time was mere 30 seconds and the data is sufficiently accurate.

One area where instrument struggles is 10 MHz ULPN OCXO at 10 Hz offset from the carrier. With specified sensitivity at -132 dBc/Hz at this point and the actual phase noise at -150 dBc/Hz it would require several thousands cross correlations (a few hours) to get close to real value.

Of course, the cost of the instrument is higher than everything else reviewed.

Another instrument worth mentioning is HA7062 Phase Noise analyzer from Holzworth Instruments. It's reasonably fast, easy to use and accurate in most cases. With every new revision though Holzworth makes trade-offs based on market demand. The "C" version is faster, but it cannot reach the floor of -190 dBc/Hz for 100 MHz ULPN OCXO. It's not important for most of their customers though.

One interesting addition to the phase noise testers is coming from Jackson Labs, but it's not released yet and we will review it once we get a hold of it.

Here's updated table.

Category

Test Instrument for testing phase noise ULPN OCXO

DCNTS

E5052B

5120A-01

APPH6000

R&S FSWP-B61

HA7062

Accuracy

*****

**** (note1)

****(note2)

***

****

****

Speed

****

* (note3)

****

*****

*****

****

Ease of use

**

****

*****

*

*****

****

Range

*****

*****

*(note4)

***

*****

****

Cost

**

*

****

*****

*

***

Ease of data retrieval

*****

*****

*

****

*****

****

Home
About NEL
について
Contact Us
Join Our Email List
Career Opportunities
XO
VCXO
TCXO
OCXO
Synthesizers
Crystals
Application Notes
Reps and Distributors
What's New
Regulatory & Compliance Statements
© NEL Frequency Controls, Inc. All rights reserved.